AppNote

Fully Automated SPE/GC/MS-Analysis of Δ9-Tetrahydrocannabinol (THC) and its Metabolites in Serum Samples


This note presents a fully automated analysis system for the determination of Δ9-tetrahydrocannabinol (THC) and its metabolites in blood serum. Automation is based on the GERSTEL MultiPurpose Sampler (MPS) equipped for solid phase extraction (GERSTEL SPE) and a module for automated eluate evaporation (GERSTEL MultiPosition Evaporation Station, mVAP).

A validated, semi-automated analysis method used for routine analysis was transferred and automated using the described system. Improvements were realized such as a reduction of the sam - ple volume used from 1 to 0.5 mL serum and the use of smaller 1 mL for mat SPE cartridges. The analysis method has been validated according to GTFCh guidelines (Society of Toxicological and Forensic Chemistry). Limits of quantification below 1 ng/mL for THC and THC-OH, extraction efficiencies between 70 and 93 % and relative standard deviations between 3.3 and 10 % were achieved. The SPE system performs sample preparation in parallel with the chromatographic run, enabling the GC/MS-system to op- erate at maximum productivity and full capacity.

Die GERSTEL benötigt die Kontaktinformationen, die Sie uns zur Verfügung stellen, um Sie bezüglich unserer Produkte und Dienstleistungen zu kontaktieren. Sie können sich jederzeit von diesen Mitteilungen abmelden. Weitere Informationen zur Abmeldung sowie zu unseren Datenschutzpraktiken und unserem Engagement zum Schutz Ihrer Privatsphäre finden Sie in unserer Datenschutzerklärung.